• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. STEM EXTENSION TAPERED CEMENTED 14 MM DIAMETER +30 MM LENGTH; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. STEM EXTENSION TAPERED CEMENTED 14 MM DIAMETER +30 MM LENGTH; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Unstable (1667)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/25/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 3007963827-2023-00321, 3007963827-2023-00322, 3007963827-2023-00323, 0002648920-2023-00285.D10-medical product: articular surface (mc) left 10 mm height, item# 42512100310, lot# 65106823.Femur cemented (cr) narrow left size 5, item# 42502005801, lot# 64957640.Tibia cemented 5 degree stemmed left size c, item# 42532006401, lot# 65032763.All-poly patella cemented 32 mm diameter, item# 42540200032, lot# 65029451.H3- customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that a patient underwent a revision of all components approximately one year nine and a half months post implantation due to instability.Attempts to obtain additional information have been made; however, no more information is available.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up is being submitted to relay additional information.No product was returned, and pictures provided were insufficient to complete visual and dimensional evaluations.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no related deviations / anomalies identified.A definitive root cause cannot be determined.The complaint is not confirmed.If any further information is received which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.H6: component code: proposed component (annex g) code is: - mechanical (g04) - stem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM EXTENSION TAPERED CEMENTED 14 MM DIAMETER +30 MM LENGTH
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18277078
MDR Text Key329856828
Report Number0001822565-2023-03333
Device Sequence Number1
Product Code JWH
UDI-Device Identifier00889024247734
UDI-Public(01)00889024247734(17)311005(10)65263755
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42557000114
Device Lot Number65263755
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/11/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/11/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10.
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexFemale
-
-