BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number D133604IL
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/25/2023
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone: (b)(6).The device was returned to biosense webster (bwi) for evaluation.A visual inspection and temperature and impedance test of the returned device were performed following bwi procedures.Visual analysis revealed that there was a reddish material inside the pebax and a hole on the surface of the tip area.The temperature and impedance test were performed, and no temperature was displayed due to an open circuit in the tip area.The hole in the pebax could be related to the issue found and the issue reported by the customer.All units are inspected prior to leaving the facility as there are functional tests and inspections at control points based on the process flow diagram (pfd), per its part number, to avoid this type of damage from leaving the facility.A manufacturing record evaluation was performed for the finished device 31033186m number, and no internal actions related to the reported complaint condition were identified.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.The instructions for use (ifu) contain the following recommendations: if the radiofrequency (rf) generator does not display temperature, verify that the appropriate cable is plugged into the rf generator.In order to prevent damage to the catheter tip, use the insertion tube supplied with the catheter to advance or retract the catheter through the hemostasis valve of the sheath.After insertion, slide the insertion tube back toward the handle.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.This product issue will be addressed through bwi's quality system.Explanation of codes: investigation findings: mechanical problem identified (c07) / investigation conclusions: unintended use error caused or contributed to event (d1102) / component code: sleeve (g04115) were selected as related to the damage to the pebax.Investigation findings: open circuit (c0205) / investigation conclusions: cause not established (d15) / component code: electrical lead/wire (g02015) were selected as related to temperature issue.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Event Description
It was reported that a patient underwent a supraventricular tachycardia (svt) with a thermocool® smart touch¿ electrophysiology catheter for which biosense webster¿s product analysis lab (pal) identified a hole in the pebax.Initially, it was reported that during the operation, there was no temperature displayed on the carto 3 system or generator.A second device was used to complete the operation.There was no adverse event reported on patient.The temperature issue was assessed as non mdr reportable.The potential risk that it could cause or contribute to a serious injury or death to the operator or patient was remote.The biosense webster, inc.Product analysis lab received the device for evaluation and per the evaluation completion on 09-nov-2023, there was a reddish material inside the pebax and a hole on the surface of the tip area.The hole in the pebax was assessed as mdr reportable.The awareness date for this reportable lab finding was 09-nov-2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18279048
MDR Text Key330021257
Report Number2029046-2023-02848
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P030031/S053
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 12/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD133604IL
Device Lot Number31033186M
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2023
Date Manufacturer Received11/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNK GENERATOR; UNK_CARTO 3
-
-