MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD VENTED AUTOFEED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD VENTED AUTOFEED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER Back to Search Results
Model Number MR290V
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2023
Event Type  malfunction  
Event Description
A distributor reported on behalf of a healthcare facility in japan, via a fisher & paykel healthcare (f&p) field representative that a mr290v vented autofeed humidification chamber cracked during patient use.There was no patient consequence reported.
 
Manufacturer Narrative
(b)(4).The complaint mr290v vented autofeed humidification chamber has been requested to be returned to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow up report upon completion of investigation.
 
Manufacturer Narrative
(b)(4).Method: the complaint mr290v vented autofeed humification chamber was returned to f&p healthcare in new zealand where it was visually inspected.A waterbath test was also conducted to confirm the location of the leak.Results: visual inspection of the returned mr290v chamber confirmed a star-shaped crack in the chamber dome just above the base.The waterbath test confirmed the leak to be at the site of the crack in the chamber.Conclusion: we are unable to determine the cause of the damage to the humification chamber.The mr290v chambers are designed and tested to conform to iso 5367 breathing tubes intended for use with anaesthetic apparatus and ventilators.Every mr290v chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber dome due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.Our user instructions that accompany the mr290v vented autofeed humidification chamber state the following: "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions, could lead to a loss of ventilation pressure." "do not use the chamber if the seals are not intact when received, or if it has been dropped.".
 
Event Description
A distributor reported on behalf of a healthcare facility in japan, via a fisher & paykel (f&p) healthcare field representative that a mr290v vented autofeed humidification chamber was leaking water after 6 days of use.On closer inspection the healthcare facility identified a crack in the chamber dome near the base.There was no patient consequence reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTED AUTOFEED HUMIDIFICATION CHAMBER
Type of Device
AUTOFEED CHAMBER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
omid taheri
17400 laguna canyon road
suite 300
irvine, CA 92618
8007923912
MDR Report Key18279685
MDR Text Key329919077
Report Number9611451-2023-01122
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K934140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR290V
Device Catalogue NumberMR290V
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received03/10/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/21/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
F&P MR810 RESPIRATORY HUMIDIFIER; F&P MR810 RESPIRATORY HUMIDIFIER; INTERSURGICAL SINGLE CIRCUIT; INTERSURGICAL SINGLE CIRCUIT; PHILLIPS TRILOGY VENTILATOR; PHILLIPS TRIOLOGY VENTILATOR; SALINE NEBULIZER
-
-