Ps437720.Method: the complaint (b)(4) adult heated wall reusable breathing circuit was received at fisher & paykel healthcare (f&p) new zealand, where visual inspection and performance testing were carried out.Results: performance testing of the device could not replicate the reported malfunction and the device was working as intended.Conclusion: we are unable to determine the cause of the reported fault as there was no fault found with the returned device.All (b)(4) adult heated wall reusable breathing circuits are visually inspected and pressure tested prior to being released for distribution.Any circuits that fail are rejected.The user instructions that accompany the (b)(4) adult heated wall reusable breathing circuit state the following: "inspect circuit before re-use, do not use if the circuit shows signs of deterioration, such as cracks, tears, or damage." "perform a pressure and leak test on the breathing system, and check for occlusions, before connecting to a patient." "do not cover the circuit with materials such as blankets, towels or bed linen." "disconnect tube by handling end connectors only, do not pull or twist tubing, as this may cause damage.".
|