LIVANOVA DEUTSCHLAND GMBH 3T HEATER COOLER; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 16-02-80 |
Device Problems
Fluid/Blood Leak (1250); Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/10/2023 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland received a report that, then the heater-cooler system 3t has been turned on, there was a massive leak of water from the area above all 4 wheels.Additionally, water runs out of the ventilation.There was no patient involvement.
|
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.G.5.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in germany.Cardioplegia block leaking water, patient pump 1 makes loud whining noises, bearing damage.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|