Please note corrections to section d9/h3; the device was returned, and h6 (device, method, results and conclusion codes).The reported event could be confirmed, since the reported "metal chips" can be confirmed.The device inspection revealed the following: the visual inspection of the nail interface at the nail adapter has shown that there are several strong deformations with some burrs are visible and that one of the prongs is slightly bent outwards.The nail adapter is in general in a very used condition, there are clearly visible stress marks at the targeting arm interface.The marks indicate that the device was previously intense used.Due to the previously described damage at the nail interface of the nail adapter it was impossible to attach a nail properly, the nail adapter is not functional as required anymore.The nail holding screws are functional as required, both devices could be inserted into the nail without any issues.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to a user related issue.The event was most likely caused by a loose connection between the nail and the adapter which did lead to a mechanical overload during use.In this relation following statement t2 alpha tibia instruments the instructions for use can be pointed out: ¿during the procedure, repeatedly check that the connections between the instruments or between implants and instruments required for the precise positioning and fixing are secure.¿ if more information is provided, the case will be reassessed.
|