SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA CORRUGATED BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C37101329-NLJ |
Device Problems
Inadequacy of Device Shape and/or Size (1583); Incomplete or Missing Packaging (2312)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/13/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the "packed anesthesia bag was not 3l".Discovered during a pre-use check.
|
|
Manufacturer Narrative
|
Other text: d4: udi number is unknown; no information has been provided to date.G5: 510k is blank, device product code is 510k exempt.D3, g1,2 email is: (b)(6).Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
Email is: regulatory.Responses@icumed.Com.One device sample was received without the original packaging.Two photos were included for evaluation; photos show a breathing bag with capacity of 1 liter instead of the 3 liters.The returned sample was reviewed and compared with the bill of materials (bom) corresponding to the part number reported and confirmed that the correct size of the breathing bag is 3 liters instead of 1 liter.The complaint was confirmed and due to manufacturing.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.The fault was considered an isolated event.The manufacturer will continue monitoring this failure condition for threshold or escalation.
|
|