Catalog Number C37101324-NLJ
Device Problems Fitting Problem (2183); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/13/2023
Event Type  malfunction  
Manufacturer Narrative
Other text: d4: udi number is unknown; no information has been provided to date.G5: 510k is blank, device product code is 510k exempt.D3, g1,2 email is: (b)(4).Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that during a pre-use check, the l-connector was unable to be connected properly.No patient injury or clinical affects was reported.
 
Manufacturer Narrative
One device was received for investigation.Per visual inspection of the device, damage was detected in the wye connector.The reported complaint is confirmed.The root cause cannot be associated with the manufacturing process since the failure reported could not be reproduced using the tools from assembly process.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
Manufacturer Contact
reed covert
MDR Report Key18283870
MDR Text Key330436043
Report Number3012307300-2023-10946
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC37101324-NLJ
Device Lot Number4393126
Was Device Available for Evaluation? No
Date Returned to Manufacturer11/21/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/09/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-