The mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the investigation of the returned device found no device deficiencies that would have contributed to "slippage." the unit has slight movement in the lock, and the index knob is loose, but when the clamp is properly positioned and put under pressure, it will not move in the locked position.This is the first time the unit has been sent in and a preventative maintenance (pm) was required.By the completion of service, the following components will have been replaced: roundhead screw, adjustment screw, label, washers, screw, nut, teflon o-ring, ball bearings, wave springs and helicoils along with general maintenance and cleaning.Additionally, the unit was sent to quality engineering for further investigation, and the findings of the service & repair report was confirmed: the lock was loose and there was little to no resistance to turn the lock between the locked and unlocked positions.Root cause - evaluation found no device deficiencies that would have contributed to the reported complaint.Probable root cause is improper or suboptimal placement of the skull clamp on the patient.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|