Catalog Number D134805 |
Device Problems
Coagulation in Device or Device Ingredient (1096); Improper or Incorrect Procedure or Method (2017)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/09/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that a patient underwent a cardiac ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and a when the catheter was removed from the patient¿s body, visual inspection revealed that that thrombus had adhered to the catheter (tip electrode, distal side).When ablating the anterior wall of the right pulmonary vein (rpv), an increase in impedance value was observed, so ablation was interrupted.The ablation was conducted again, but there was no improvement on impedance value.The catheter was removed from the patient's body, and visual inspection revealed that thrombus had adhered to the catheter.The catheter was replaced to complete the procedure.No adverse patient consequence was reported.Additional information was received, and it was reported that when the thrombus occurred, irrigation flow was kept low flow due to wrong setting.As the generator setting was for 4mm ablation catheter, it remained ¿low¿ irrigation flow during ablation.Heparinized normal saline was used.The rise in impedance is not mdr reportable as the change in impedance was not reported to exceed any cut off value and no error was displayed regarding impedance.The thrombus issue is mdr reportable.
|
|
Manufacturer Narrative
|
E1.Initial reporter phone:(b)(6).If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|
|
Manufacturer Narrative
|
The biosense webster, inc.Product analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
|
Search Alerts/Recalls
|
|