BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number D134805
Device Problems Coagulation in Device or Device Ingredient (1096); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/09/2023
Event Type  malfunction  
Event Description
It was reported that a patient underwent a cardiac ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and a when the catheter was removed from the patient¿s body, visual inspection revealed that that thrombus had adhered to the catheter (tip electrode, distal side).When ablating the anterior wall of the right pulmonary vein (rpv), an increase in impedance value was observed, so ablation was interrupted.The ablation was conducted again, but there was no improvement on impedance value.The catheter was removed from the patient's body, and visual inspection revealed that thrombus had adhered to the catheter.The catheter was replaced to complete the procedure.No adverse patient consequence was reported.Additional information was received, and it was reported that when the thrombus occurred, irrigation flow was kept low flow due to wrong setting.As the generator setting was for 4mm ablation catheter, it remained ¿low¿ irrigation flow during ablation.Heparinized normal saline was used.The rise in impedance is not mdr reportable as the change in impedance was not reported to exceed any cut off value and no error was displayed regarding impedance.The thrombus issue is mdr reportable.
 
Manufacturer Narrative
E1.Initial reporter phone:(b)(6).If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Manufacturer Narrative
The biosense webster, inc.Product analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Manufacturer Narrative
It was reported that a patient underwent a cardiac ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and a when the catheter was removed from the patient¿s body, visual inspection revealed that that thrombus had adhered to the catheter (tip electrode, distal side).When ablating the anterior wall of the right pulmonary vein (rpv), an increase in impedance value was observed, so ablation was interrupted.The ablation was conducted again, but there was no improvement on impedance value.The catheter was removed from the patient's body, and visual inspection revealed that thrombus had adhered to the catheter.The catheter was replaced to complete the procedure.No adverse patient consequence was reported.Additional information was received, and it was reported that when the thrombus occurred, irrigation flow was kept low flow due to wrong setting.As the generator setting was for 4mm ablation catheter, it remained ¿low¿ irrigation flow during ablation.Heparinized normal saline was used.Device evaluation details: the device was returned to biosense webster (bwi) for evaluation.Visual inspection, temperature, impedance, pump, and pressure gauge test of the returned device were performed following bwi procedures.Visual analysis revealed no damage or anomalies on the device.No char, thrombus or clot residues were observed during the inspection.The temperature and impedance test were performed and the device was found working correctly.No temperature or impedance issues were observed.A pump and pressure gauge test were performed, and the device was irrigating correctly.No irrigation issues were observed.A manufacturing record evaluation was performed for the finished device 31134490l number, and no internal action related to the complaint was found during the review.Additionally, the manufacture and expiration dates have been provided.Therefore, fields d4.Expiration date and h4.Device manufacture date have been populated.The potential cause of the thrombus or clot residues observed by the customer could be related to the low flow during the irrigation, due to wrong setting.Even when no thrombus residues were observed, the wrong settings could be related to the issue reported by the customer.Therefore, the customer complaint was confirmed.The instructions for use contain the following recommendations: if the radiofrequency (rf) generator does not display temperature, verify that the appropriate cable is plugged into the rf generator.Flush the catheter with heparinized saline before insertion into the body.Always maintain a constant heparinized normal saline infusion to prevent coagulation within the lumen of the catheter.Do not use the catheter without irrigation flow.The temperature sensor is used to verify that the irrigation flow rate is adequate.Before initiating the application of rf energy, a decrease in electrode temperature confirms the onset of saline irrigation of the ablation electrode.Monitoring the temperature from the electrode during the application of rf energy ensures that the irrigation flow rate is being maintained.Using tip temperature to guide ablation could result in deeper lesions and increased risk for collateral damage.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18288489
MDR Text Key330435354
Report Number2029046-2023-02879
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD134805
Device Lot Number31134490L
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received04/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/12/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SMARTABLATE GEN. KIT (JAPAN).; SMARTABLATE PUMP KIT (JAPAN).; UNK_CARTO 3.
-
-