U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLSTENT ENTERAL ENDOPROSTHESIS; STENT, COLONIC, METALLIC, EXPANDABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLSTENT ENTERAL ENDOPROSTHESIS; STENT, COLONIC, METALLIC, EXPANDABLE Back to Search Results
Model Number M00565570
Device Problems Positioning Failure (1158); Material Puncture/Hole (1504); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2023
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that a wallstent enteral uncovered stent was to be implanted in the duodenum to treat a malignant stricture secondary to duodenal cancer during an intestinal stent implantation procedure performed on (b)(6) 2023.During the procedure, the stent could not be deployed inside the patient because there was resistance at the tip of the delivery system.The device was removed from the patient, and it was noted that a stent wire had punctured the tip of the outer sheath.Another wallstent enteral stent was used to complete the procedure.There were no patient complications reported as a result of this event, and the patient's condition after the procedure was reported to be stable.
 
Manufacturer Narrative
Block h6: imdrf device code a041001 captures the reportable event of a stent wire that punctured the sheath.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLSTENT ENTERAL ENDOPROSTHESIS
Type of Device
STENT, COLONIC, METALLIC, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18288589
MDR Text Key330462873
Report Number3005099803-2023-06514
Device Sequence Number1
Product Code MQR
UDI-Device Identifier08714729285069
UDI-Public08714729285069
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K000281
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00565570
Device Catalogue Number6557
Device Lot Number0031071348
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexFemale
Patient Weight55 KG
-
-