• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® DRYSEAL FLEX INTRODUCER SHEATH; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® DRYSEAL FLEX INTRODUCER SHEATH; INTRODUCER, CATHETER Back to Search Results
Catalog Number DSF2033
Device Problems Difficult to Insert (1316); Patient-Device Incompatibility (2682)
Patient Problem Vascular Dissection (3160)
Event Date 11/13/2023
Event Type  Injury  
Manufacturer Narrative
According to the gore® dryseal flex introducer sheath instructions for use, adverse events that may occur and / or require intervention include vascular trauma (i.E., dissection, rupture, perforation, tear, etc.).W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
The following information was reported to gore: on (b)(6) 2023, this patient underwent an emergency endovascular treatment for a ruptured thoracic aortic aneurysm using gore® dryseal flex introducer sheath.The 20 fr sheath was inserted from the left side of the patient.The angiography after stent graft placement confirmed the left internal iliac artery occlusion.No treatment was performed and the procedure was completed.It was reported that the origin of the left external iliac artery was narrow and noted insertion difficulty of the sheath even after percutaneous transluminal angioplasty was performed.Although the access angiography did not show any issues, it was suspected that the left internal iliac artery was occluded due to dissection from the left external iliac artery to the left common iliac artery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® DRYSEAL FLEX INTRODUCER SHEATH
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 1 B/P
32360 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
monique nwanze
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key18296652
MDR Text Key330115267
Report Number3007284313-2023-02931
Device Sequence Number1
Product Code DYB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K160254
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue NumberDSF2033
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age79 YR
Patient SexFemale
-
-