• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number A2025-200
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2023
Event Type  malfunction  
Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event Description
It was reported that the procedure was to treat a 70% stenosed lesion in the anterior tibial artery.The 2.5x200mm armada balloon dilatation catheter (bdc) was used in the procedure; however, the luer connection between the bdc and the inflation device was not aligning properly.There was a leak from the hub and the armada balloon would only hold pressure by firmly holding the inflation device onto the bdc hub during inflation.The same inflation device was used with other balloons in the procedure without any issues.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
 
Manufacturer Narrative
Visual and functional inspections were performed.The reported loose or intermittent connection issue could not be confirmed during return analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported complaint.There was no damage noted to the device during the inspection prior to use, which suggests a product quality issue did not contribute to the reported difficulties.In this case, it is possible that the luerlock of the inflation device was not properly connected to the hub or possibly over torqued resulting in a loose/intermittent connection, however a definitive cause for the reported difficulties cannot be determined since the complaint could not be confirmed during return analysis.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 14 PTA CATHETER
Type of Device
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18297587
MDR Text Key330133868
Report Number2024168-2023-13742
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
K102705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA2025-200
Device Lot Number3041741
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-