COCHLEAR LIMITED NUCLEUS CI622 COCHLEAR IMPLANT WITH SLIM STRAIGHT ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number CI632
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem Vertigo (2134)
Event Date 03/22/2023
Event Type  Injury  
Manufacturer Narrative
This report is submitted on dec 11, 2023.
 
Event Description
Per the audiologist, the patient experienced vertigo which was determined to be device related and a decline in hearing.The vestibular assessment showed damage to the vestibular nerve.The cochlear implant was evaluated on (b)(6) 2023 using the integrity test system.The results indicated neither evidence of malfunction of the receiver/stimulator nor electrode faults.The vertigo was treated with a medication.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI622 COCHLEAR IMPLANT WITH SLIM STRAIGHT ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LIMITED
1 university avenue
macquarie university, 2109
AS  2109
Manufacturer (Section G)
COCHLEAR LIMITED
1 university avenue
macquarie university
nsw 2109
AS   2109
Manufacturer Contact
roberta loveday
1 university avenue
macquarie university, nsw 2109
AS   2109
MDR Report Key18298603
MDR Text Key330110314
Report Number6000034-2023-03838
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502036573
UDI-Public(01)09321502036573(11)220621(17)240620
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberCI632
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/21/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age83 YR
Patient SexFemale
-
-