• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL PROCLAIM¿ DRG IMPLANTABLE PULSE GENERATOR; DRG IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL PROCLAIM¿ DRG IMPLANTABLE PULSE GENERATOR; DRG IPG Back to Search Results
Model Number 3664
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Implant Pain (4561)
Event Date 11/14/2023
Event Type  Injury  
Event Description
It was reported that the patient experienced pain at the ipg site.As a result, surgical intervention was undertaken on (b)(6) 2023 where in the ipg was repositioned to address the issue.
 
Manufacturer Narrative
Date of event is estimated.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM¿ DRG IMPLANTABLE PULSE GENERATOR
Type of Device
DRG IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
heidi syndergaard
6901 preston road
plano, TX 75024
9723098000
MDR Report Key18300402
MDR Text Key330161034
Report Number1627487-2023-05597
Device Sequence Number1
Product Code PMP
UDI-Device Identifier05415067020215
UDI-Public05415067020215
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number3664
Device Lot NumberT00005066
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/14/2023
Initial Date FDA Received12/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/05/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DRG LEAD (X2)
Patient Outcome(s) Other;
Patient SexFemale
Patient Weight109 KG
-
-