Catalog Number 130790036
Device Problems Fracture (1260); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2023
Event Type  malfunction  
Event Description
Locking metaglene screw head detached from the screw and cannot be locked with base plate during surgery.It was tried on patient but removed once defect observed.The screw was not implanted on patient, surgery continued with another screw.There was no surgical delay.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint #
=
> (b)(4) investigation summary
=
> according to the information received, locking metaglene screw head detached from the screw and cannot be locked with base plate during surgery.It was tried on patient but removed once defect observed.The screw was not implanted on patient, surgery continued with another screw.The device associated with this report was returned to depuy synthes for evaluation.Visual inspection of the returned sample revealed that no defects were observed in the threads or the tip of the dxtend screw lock d4.5x36mm.The screw was evaluated, and no defects or damage were observed.The screw can be easily screwed and unscrewed in the head of the screw lock.A dimensional inspection was performed for the screw lock and met specifications.A functional test was unable to be performed due to the locking screwdriver not returned.It was not possible to verify if there was any problem with the screwdriver attached to the screw.The complaint condition was not able to be replicated.The overall complaint was unconfirmed as the observed condition of the dxtend screw lock d4.5x36mm would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing: dwg-8e070132 feature: screw head diameter specification: 4.00 -0.10 mm measured dimension: 3.95 mm device used: mitutoyo digimatic caliper a20276546 id# cd78619 drawing: dwg-8e070128 feature: length screw specification: 36.00 +/- 0.20 mm measured dimension: 36.10 mm device used: mitutoyo digimatic caliper a20276546 id# cd78619 device history lot
=
> the dhr analysis returned shows an initial conformance of this product with regards to its specification.For this batch, there was no deviation or non-conformance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND SCREW LOCK D4.5X36MM
Type of Device
DELTA XTEND IMPLANTS : SCREWS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18300750
MDR Text Key330164801
Report Number1818910-2023-25083
Device Sequence Number1
Product Code HSD
UDI-Device Identifier10603295027850
UDI-Public10603295027850
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
K192855
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number130790036
Device Lot Number5551916
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/16/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-