Model Number N/A
Device Problems Difficult to Insert (1316); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/27/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).G2: foreign: austria.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the g7 inlay did not stick in the shell as the rim was damaged.There was no patient impact or delay.There is no additional information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: updated: b4.D4 expiration and udi.G3.G6.H2.H4.H6.H10.Visual examination of the provided pictures identified frayed pieces of locking feature, with a cut in the scalloped face.A review of the device history records identified no related deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
There is no update to the prior event description provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL E1 LINER 36MM F
Type of Device
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18303136
MDR Text Key330260974
Report Number0001825034-2023-02946
Device Sequence Number1
Product Code PBI
UDI-Device Identifier00880304526433
UDI-Public(01)00880304526433(17)280411(10)7491852
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number010000858
Device Lot Number7491852
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/09/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
UNKNOWN SHELL
Patient SexPrefer Not To Disclose
-
-