SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID
|
Back to Search Results |
|
Catalog Number HL-390 |
Device Problems
Device Alarm System (1012); Failure to Sense (1559)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/01/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the device was not detecting disposables and continued alarming.The reporter tried to reproduce the fault, but he could not duplicate it.The unit was being tested, and the product has been working properly.The failure occurred during preventive maintenance test.No injury.
|
|
Manufacturer Narrative
|
B3: date of event is unknown, no information has been provided to date.Date entered is our best estimate.H3 other: device has not been returned for investigation to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
H6 - updated.No device was received for investigation.Therefore, we are unable to confirm the reported complaint.If we receive the device, we will reopen the investigation for further evaluation.The most probable cause is an inoperable disposable switch sensor.A review of the service history found could not be performed with serial number provided.
|
|