MAUDE Adverse Event Report: BIOSENSE WEBSTER INC DECANAV ELECTROPHYSIOLOGY CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC DECANAV ELECTROPHYSIOLOGY CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Catalog Number R7F282CT
Device Problems Communication or Transmission Problem (2896); Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/13/2023
Event Type  malfunction  
Manufacturer Narrative
The biosense webster, inc.Product analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Event Description
It was reported that a patient underwent an atrial flutter (afl) ablation procedure with a decanav electrophysiology catheter and the carto 3 system was displaying an error 7,"current leakage error", and there was noise on the decanav, as well as on the body surface signals.The noise was also on the recording system.The cable was replaced without resolution.When the catheter was replaced, the issue was resolved, and the procedure was continued.No adverse patient consequence was reported.Initially, this event was assessed as not mdr reportable.However, on 15-nov-2023, additional information was received stating that there were no intact electrocardiogram (ecg) signals while the decanav was connected and while the catheter was in the coronary sinus (cs).When the bad catheter was disconnected, the ecg signals returned to both carto and recording system.A new cable did not fix this issue and so the catheter was then replaced.The additional information received makes this event mdr reportable product malfunction with an awareness date of 15-nov-2023.
 
Manufacturer Narrative
The device evaluation was completed on 05-jan-2024.It was reported that a patient underwent an atrial flutter (afl) ablation procedure with a decanav electrophysiology catheter and the carto 3 system was displaying an error 7,"current leakage error", and there was noise on the decanav, as well as on the body surface signals.The noise was also on the recording system.The cable was replaced without resolution.When the catheter was replaced, the issue was resolved, and the procedure was continued.No adverse patient consequence was reported.Initially, this event was assessed as not mdr reportable.However, on 15-nov-2023, additional information was received stating that there were no intact electrocardiogram (ecg) signals while the decanav was connected and while the catheter was in the coronary sinus (cs).When the bad catheter was disconnected, the ecg signals returned to both carto and recording system.A new cable did not fix this issue and so the catheter was then replaced.Device evaluation details: the device was returned to biosense webster (bwi) for evaluation.A visual inspection, magnetic sensor functionality, and electrical test of the returned device were performed following bwi procedures.Visual analysis revealed no damage or anomalies on the device.The device was connected to the carto 3 system, and it was recognized and visualized correctly.No issues or errors were observed.Afterward, an electrical test was performed, and no noise issues were found.A manufacturing record evaluation (mre) was performed for the finished device 31138998m number, and no internal actions related to the reported complaint condition were identified.Based on the completed mre, the manufactured date and expiration date have been provided.Therefore, fields d4.Expiration date and h4.Device manufacture date have been populated with appropriate information.The issue reported by the customer could not be replicated during the product investigation.Other issues or circumstances may have occurred during the usage of the device that compromised its performance.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DECANAV ELECTROPHYSIOLOGY CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18314850
MDR Text Key330466616
Report Number2029046-2023-02925
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberR7F282CT
Device Lot Number31138998M
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2023
Is the Reporter a Health Professional? No
Date Manufacturer Received01/05/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/05/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNK BRAND CABLE.; UNK BRAND CABLE.; UNK_CARTO 3.
-
-