OLYMPUS WINTER & IBE GMBH VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 0°, AUTOCLAVABLE; ENDOEYE HD II (WA50040A)
|
Back to Search Results |
|
Model Number WA50040A |
Device Problem
Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/28/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation, and the customer¿s allegation was not confirmed as the image issue could not be reproduced.The following non-reportable malfunctions were found during the device evaluation: the video connector has a loose serial ring, the outer tube had a dent, and a white dot was found on the image.The investigation is ongoing and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Event Description
|
The customer reported to olympus that the video telescope "endoeye hd ii", 10 mm, 0°, autoclavable had the image tone become green when the endoscope is connected or the device starts.The issue was found during preparation for use.There were no reports of patient harm.Related patient identifiers are as follows: c23519642.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the likely cause of the reported event is due to wrong handling by the user or, wear and tear.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|