Model Number CS100 |
Device Problems
Break (1069); Structural Problem (2506)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/24/2023 |
Event Type
malfunction
|
Event Description
|
The china customer reported the lower door of the coverstainer is broken.A picture provided shows the top left corner of the door has detached from the rest of the door.The issue occurred when putting down the lower door after replacing reagents.There was no harm to the user.The instrument is still able to be used to complete testing.Diagnostics were not altered.There was no direct or indirect patient harm or user harm reported.The investigation is ongoing, captured in capa01211, and a supplemental will be provided when new pertinent information is received.
|
|
Manufacturer Narrative
|
No erroneous staining result was reported by the customer in connection with this incident.No patient or user harm was indicated.A1-a6: patient information has not been provided by the user.While there was no direct, or indirect, user harm reported for this event; it is being reported due to the potential for harm if the event were to reoccur.The instrument experienced a similar failure mode related to reportable complaint with injury, reference number mdr 3003423869-2023-00084, (b)(4).Therefore, this report is being filed as part of agilent's commitment to due diligence reporting.
|
|
Manufacturer Narrative
|
Agilent technologies has initiated a recall in april 2024 with the scope of notifying all coverstainer customers of the following issue: the lower door that opens to the reagent bottles is cracking and/or breaking off in the upper corner of the plexiglass.Capa01211 has addressed the issue and the investigation has been completed.Agilent technologies has determined the root cause as such; the lower door support bracket, which connects the hinges to the lower door, is only attached to the plexiglass and not to the metal frame which is the main support structure of the lower door.The weight of the plexiglass door is in turn cracking/breaking at this joint.Customers have been notified of this issue via a customer letter.Agilent technologies worked with the manufacturer to re-design the support bracket for the lower door hinges, so the weight of the lower door assembly is not supported by the plexiglass, but by the metal frame of the lower door.This new bracket is being verified for effectiveness; once completed this bracket will be installed at the customer's next service visit.
|
|
Event Description
|
The china customer reported the lower door of the coverstainer is broken.A picture provided shows the top left corner of the door has detached from the rest of the door.The issue occurred when putting down the lower door after replacing reagents.There was no harm to the user.The instrument is still able to be used to complete testing.The field service engineer (fse) serviced the instrument and replaced the lower door which resolved this malfunction.The instrument is fully operational, within specification, and ready for use.Diagnostics were not altered.There was no direct or indirect patient harm or user harm reported.The investigation is ongoing, captured in capa01211, and a supplemental will be provided when new pertinent information is received.
|
|
Manufacturer Narrative
|
The following fields were updated.B4, b5, g3, g6, h2, h3, h6, h7, h11.
|
|
Search Alerts/Recalls
|
|