MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR ESSENTI
|
Back to Search Results |
|
Model Number 37612 |
Device Problems
Failure to Power Up (1476); Delayed Charge Time (2586); Charging Problem (2892); Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/05/2023 |
Event Type
malfunction
|
Event Description
|
Information was received from a patient's representative regarding an external device.The reason for call was caller states that the recharger works intermittently and sometimes it doesn't work at all.Caller states the patient will turn on the charger and it will not turn on and she will have to wait and keeping trying before they can get it to turn on.The caller stated that the recharger does not progress 50% charged , and stated " it only charges it to 50% capacity".The caller also stated that the patient's implantable neurostimulator (ins) was also taking longer to charge.And is also not getting past 50%.The caller confirmed that there is no damage to the desktop charger and that the green light is on.Patient services (pss) had the caller plug the dtc into the recharger and confirmed that the recharge was charging and state that it was blinking about half.The caller stated that the patient was not with them at the time of the call and would call ps back to perform additional troubleshooting steps ( check coupling boxes ).Patient is switching over to wireless recharger (wr).No symptoms reported.
|
|
Manufacturer Narrative
|
Continuation of d10: product id 37751 lot# serial# (b)(6) product type recharger section d information references the main component of the system.Other relevant device(s) are: product id: 37751, serial/lot #: (b)(6).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from the manufacturer¿s representative (rep) reported the patient was upgraded to a new recharger, but it was unclear why the older charger was having issues.
|
|
Manufacturer Narrative
|
H3.Analysis of the recharger (s/n (b)(6) found the device tested okay.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|