Event Description
It was reported that during a bench test the device had no flow.Also the temperature did not go past 35 on temperature check while testing the device.Display not going up to operating temperature.No alarms.Tried the overtemp and machine turned off as it should.However, when trying to turn it back on there was no re-circulation of fluid but had no alarms.
 
Manufacturer Narrative
Other text: b3: date of event is unknown, no information has been provided to date.D3, g1, and g2 email is: (b)(6).Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 FAST FLOW FLUID WARMER
Type of Device
WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
reed covert
MDR Report Key18321031
MDR Text Key330467243
Report Number3012307300-2023-11020
Device Sequence Number1
Product Code BSB
UDI-Device Identifier50695085829506
UDI-Public50695085829506
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK020043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8002950
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/09/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-