On the day of activation of the reactiv8 system post-ipg relocation surgery, the clinical specialist found that all electrodes on the left and one on the right lead had an out-of-range/high impedance.The surgeon decided to perform a second surgery to replace both leads.Both leads were removed intact, and new leads were implanted with no report of patient harm or injury.The same ipg was used.The removed leads were inadvertently discarded at the hospital.There was no part analysis performed.The manufacturing record was reviewed- no non-conformance was found.The device was activated two weeks later with no issues.The patient can now run bilateral stimulation.The repositioning of the (ipg) can be found under reference manufacturing report number 3013017877-2023-00054.
|