Information was received from a healthcare provider regarding a patient who was implanted with an implantable neurostimulator (ins) for gastroparesis.They reported that during the revision surgery, after the new battery was placed, impedance testing showed out of specification.At that point it was determined that the leads were likely completely malpositioned.They decided to remove the existing leads and place a new stimulator with a new set of leads in the left abdomen.
|
See manufacturer¿s report #3004209178-2023-07193 for concomitant ins information.Continuation of d10: product id 3116, serial# (b)(6), implanted: (b)(6) 2023.Product type implantable neurostimulator product id 435135, serial# (b)(6), implanted: (b)(6) 2002, explanted: (b)(6) 2023.Product type lead product id 435135, serial# (b)(6), implanted: (b)(6) 2002, explanted: (b)(6) 2023.Product type lead section d information references the main component of the system.Other relevant device(s) are: product id: 435135, serial/lot #: (b)(6), ubd: 15-aug-2004, udi#: (b)(4); product id: 435135, serial/lot #: (b)(6), ubd: 29-aug-2004, udi#: (b)(4) h3:analysis of the lead (s/n (b)(6)) revealed a breach of the outer insulation due to environmentally assisted degradation in the body of the lead.Analysis of the lead (s/n (b)(6)) revealed a breach of the outer insulation due to environmentally assisted degradation in the body of the lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|