• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065753057
Device Problems Energy Output Problem (1431); Suction Problem (2170)
Patient Problems Cusp Tear (2656); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 11/18/2023
Event Type  Injury  
Event Description
A nurse reported that during a cataract surgery an ophthalmic operating console failed to perform the fill test system was restarted and the test was performed.However, during the surgery, the system experienced a vacuum delivery failure, resulting in insufficient aspiration.The surgeon attempted various measures such as changing the handpiece, tip, and cassette, but the issue persisted, leading to the system being turned off multiple times.Despite reconnecting the cassette and altering the handpiece, the vacuum problem persisted, causing an increase in ultrasound impulse that led to the rupture of the zonules and displacement of the crystalline towards the vitreous in the patient's eye.To mitigate the situation, the surgery proceeded using an alternative ophthalmic system, necessitating a posterior vitrectomy, posterior planectomy, and sulcus lens implant to complete the procedure additional information has been requested none received till date.This report pertains to ophthalmic operating system involved in this reported event.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
Specific product identifiers (lot number, batch number, and/or serial number) were not provided and could not be determined at this time.However, all device history records are reviewed prior to product release to ensure the product was manufactured in compliance with the device master record and meets release criteria.A review for complaints reported against this serial number cannot be performed as the serial number is unknown.The serial is unknown.Therefore, a service history review cannot be performed.Posterior capsule tear is an issue that is occasionally reported with cataract surgery.However, a review of the complaint trends shows that the frequency reported is within known levels for this event.A root cause cannot be determined conclusively.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Event Description
Additional information received indicated that surgery was not possible to perform.Patient was referred to referred to retinology who was suggested phacoemulsification surgery and insertion of an artisan type lens.Patient required a reoperation after the surgery.
 
Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18324963
MDR Text Key330485719
Report Number2028159-2023-01676
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657530571
UDI-Public00380657530571
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K161794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup,Followup
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065753057
Device Lot Number13VKCU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CENTURION SURGICAL PROCEDURE PAK; HANDPIECE
Patient Outcome(s) Required Intervention; Other;
Patient Age64 YR
Patient SexFemale
-
-