The event involved a 183 cm (72") pur set d'infusione, perf.Con valvola, 4 accessi con bcv, camera, filtro 15¿, rub.Con clave¿, ll gir.The customer reported mismatch of one of the routes of the chemotherapy tubing between the 'var' and the tubing, causing a leak of the chemotherapy product to the ground and requiring the repreparation of the bag for the patient.The event was observed during use on a patient, at the beginning of the administration.There was a delay in therapy due to the need to redo the chemo bag.The drug administrated was carboplatin.The patient received the full intended dose, but with the new bag.There was no undesirable clinical consequences, and nobody was hurt as a consequence of this event.
|