On (b)(6), 2023, a regent mechanical heart valve 21agfn-756 (serial: (b)(6) was used in an aortic valve replacement surgery.During the process of implanting the valve, it was a crack was discovered in the valve frame, so the valve was removed and used another valve 21agfn-756 (serial: (b)(6) to complete the operation.The valve was rotated with the valve rotator, no resistance while rotating, no instruments encountered the valve at the time of fracture, holder handle was fully inserted into orifice at 90 degree angle, valve was in closed position when it was rotated.The patient remained on cardiopulmonary bypass throughout the procedure.There was a 20 minute delay in the procedure that was considered clinically acceptable by the physician.The patient remained hemodynamically stable throughout the procedure.The patient's condition is currently stable and discharged.
|
An event of the orifice being found to be cracked after the device was implanted was reported.The device was received for examination and the orifice crack was confirmed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|