Model Number 866389 |
Device Problem
Wireless Communication Problem (3283)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/22/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
|
|
Event Description
|
The customer reported that there was a loss of central monitoring on the 2nd floor, 3rd floor and emergency department.It is unknown if the device was in use at the time of the event.There was no adverse event reported.
|
|
Manufacturer Narrative
|
Review of the provided information indicates that the device was in use at the time of the event.A philips technical consultant (tc) was dispatched for onsite service.Results of functional testing indicate the core b switch was defective.The cause of the reported problem was bad switch.The reported problem was confirmed.A replacement switch is being shipped to the customer to resolve the issue.If additional information is received the complaint file will be reopened.
|
|