Catalog Number 2R8486 |
Device Problems
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/24/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
E1: initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a clearlink non-dehp solution set had a hole about 1.5 inches long in the middle of the tubing.This was observed during priming.The total parenteral nutrition (tpn) containing tube leaked though the clamp and cap was on the set.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the actual device was provided for evaluation.A visual inspection with the naked eye observed a cut in the tubing.The reported condition was verified.The cause of the condition was a manufacturing related issue.It was determined the set was placed incorrectly in the pouch prior to packaging, leaving parts of the set outside the pouch and exposed to the packaging machine blades causing the cut in the tubing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Search Alerts/Recalls
|