Model Number PM1272
Device Problems Communication or Transmission Problem (2896); Interrogation Problem (4017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/11/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented to the hospital for a scheduled implant procedure.Prior to the procedure, it was noted that the pacemaker exhibited interrogation problem and was unable to be programmed.The physician elected to replace the pacemaker and proceeded with the procedure.The patient was discharged post-surgery.
 
Manufacturer Narrative
The reported events of interrogation, communication and transmission problems were not confirmed.As received, a complete device was returned with a battery voltage below the elective replacement indicator level.The analysis revealed that the electrical and mechanical conditions were normal and no interrogation anomalies were observed.Also, a longevity assessment was performed, and it indicates that the device was in the normal range of operation with appropriate remaining longevity.No anomalies were seen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY MRI
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18332702
MDR Text Key330584849
Report Number2017865-2023-94852
Device Sequence Number1
Product Code LWP
UDI-Device Identifier05414734509572
UDI-Public05414734509572
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPM1272
Device Lot NumberP000174908
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-