The customer reported the patient's feeding tube was found 10 cm out from the previously documented position, so the tube was reinserted and an x-ray was done to confirm the position.On the x-ray it was found that the feeding tube had fractured in the midportion of the tube.The tip of the proximal tube was in the esophagus, and the distal half of the tube was in the stomach.The distal portion was retrieved from the patient¿s stomach.
|
The device history record (dhr) review could not be performed because no lot number was provided.One sample was received.Only the part of the tube up to the number 25 was received; the tip was not received.The sample presented a burst, the reported condition is confirmed, however the issue was not caused in the manufacturing plant.A walk through was performed of the manufacturing process.All process and controls were found properly followed.Production personnel perform a 100% visual inspection during the packaging process, to detect and discard any identified issues.The root cause could not be determined with the information available.No further actions are planned.This complaint will be used for tracking and trending purposes.
|