RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 V60PLUS VENTILATOR
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2023
Event Type  malfunction  
Event Description
Philips received a complaint from customer biomed, reporting the v60 ventilator would not go into standby mode.The device was not in clinical use.There was no report of harm, and no impact to patient, nor user.A philips remote service engineer (rse) evaluated the issue with the biomed engineer (bme) and reported the average air standard liters per minute (slpm) readout on the pneumatics screen was -2.5.The rse informed the bme the reading was out of tolerance and advised full performance verification testing to diagnose the issue.
 
Manufacturer Narrative
H10: air flow sensor assembly was recommended for replacement.The customer confirmed the part has not been made available for device repair.The repair for this unit cannot be conducted at this time due to the part(s) being on backorder.The material has already been requested and an order for the part(s) was placed to conduct the repair of this unit.The material ordered (air flow sensor assembly) aligns with the recommended repair of the alleged malfunction per the service manual.When the parts become available the repairs will be conducted.If new information is received and suggest that the device has additional malfunctions, the record will be reopened, and an investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18334547
MDR Text Key330596491
Report Number2518422-2023-35820
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/01/2024
Date Device Manufactured09/13/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-