RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 V60PLUS VENTILATOR |
Device Problem
Device Sensing Problem (2917)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/13/2023 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint from customer biomed, reporting the v60 ventilator would not go into standby mode.The device was not in clinical use.There was no report of harm, and no impact to patient, nor user.A philips remote service engineer (rse) evaluated the issue with the biomed engineer (bme) and reported the average air standard liters per minute (slpm) readout on the pneumatics screen was -2.5.The rse informed the bme the reading was out of tolerance and advised full performance verification testing to diagnose the issue.
|
|
Manufacturer Narrative
|
H10: air flow sensor assembly was recommended for replacement.The customer confirmed the part has not been made available for device repair.The repair for this unit cannot be conducted at this time due to the part(s) being on backorder.The material has already been requested and an order for the part(s) was placed to conduct the repair of this unit.The material ordered (air flow sensor assembly) aligns with the recommended repair of the alleged malfunction per the service manual.When the parts become available the repairs will be conducted.If new information is received and suggest that the device has additional malfunctions, the record will be reopened, and an investigation will be performed.
|
|