ZIMMER BIOMET, INC. G7 STR MONOBLOCK SHELL INSRTR; INSTRUMENT, HIP Back to Search Results
Catalog Number 110003450
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/20/2023
Event Type  malfunction  
Event Description
It was reported that while preparing for the case, the threads on the cup inserter were found to be damaged and would not work.Another inserter was used to complete the case.Attempts have been made and no further information is available.
 
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
Cmp-(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected upon visual inspection the shaft of the device shows scuffing along with indentations on the strike plate.The tip of the device is damaged no further analysis was done.The complaint is confirmed based on the evaluation of the returned product.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR MONOBLOCK SHELL INSRTR
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18334649
MDR Text Key330597449
Report Number0001825034-2023-02928
Device Sequence Number1
Product Code PBI
UDI-Device Identifier00887868462481
UDI-Public(01)00887868462481(11)230619(10)66063582
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number110003450
Device Lot Number66063582
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/19/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-