Catalog Number 110003450 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/20/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that while preparing for the case, the threads on the cup inserter were found to be damaged and would not work.Another inserter was used to complete the case.Attempts have been made and no further information is available.
|
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
Cmp-(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected upon visual inspection the shaft of the device shows scuffing along with indentations on the strike plate.The tip of the device is damaged no further analysis was done.The complaint is confirmed based on the evaluation of the returned product.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information to report at this time.
|
|