OLYMPUS WINTER & IBE GMBH VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE; LAPAROSCOPE, GYNECOLOGIC Back to Search Results
Model Number WA50042A
Device Problems No Display/Image (1183); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation, and the customer¿s allegation was confirmed.The evaluation found the following reportable issues: the charged couple device was damaged causing the purple image and green stripe and the e111 fog-free error.No other issues were found at evaluation.The investigation is ongoing and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus that the video telescope "endoeye hd ii", 10 mm, 30°, autoclavable had a purple image.The issue was found during preparation for use.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE
Type of Device
LAPAROSCOPE, GYNECOLOGIC
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18335757
MDR Text Key330607180
Report Number9610773-2023-03640
Device Sequence Number1
Product Code HET
UDI-Device Identifier04042761074971
UDI-Public04042761074971
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
K111788
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA50042A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2023
Was the Report Sent to FDA? No
Date Manufacturer Received11/28/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-