Model Number 866389
Device Problems Alarm Not Visible (1022); Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2023
Event Type  malfunction  
Event Description
The customer reported that alarms from the monitors were not displayed at the control center.The device was in use on patient at time of event, there was no adverse event reported.
 
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.E1: reporter institution phone number (b)(6).E1: reporter phone number (b)(6).
 
Manufacturer Narrative
Philips received a complaint on the patient information center ix, indicating that the alarms from the monitors were not displayed on the control panel.A good faith effort (gfe) was performed to further clarify the allegation, and it was provided that the issue was specifically related to the unexpected shutdown.The device was in use on patient at time of event, there was no adverse event reported.The remote service engineer (rse) spoke to the customer and found that there was a control center failure and restart.The rse determined that the control panel shut down unexpectedly during operation and explained that in most cases, this is a result of problems with the power supply.The rse was unable to find any entries about a hardware problem with the pc itself.Based on the information available and the testing conducted, the cause of the reported problem is unknown.The reported problem was not confirmed.The rse advised the customer to check the wiring of the power supply as well as the ups.It is unknown what caused the issue, as the customer did not call back for further assistance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATIENT INFORMATION CENTER IX
Type of Device
PATIENT INFORMATION CENTER IX
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18336139
MDR Text Key330609990
Report Number1218950-2023-00931
Device Sequence Number1
Product Code MHX
UDI-Device Identifier00884838093041
UDI-Public00884838093041
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K183387
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866389
Device Catalogue Number866389
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/05/2024
Date Device Manufactured02/24/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-