Catalog Number 7236-2-852
Device Problem Device Dislodged or Dislocated (2923)
Patient Problems Fall (1848); Joint Dislocation (2374)
Event Date 11/18/2023
Event Type  Injury  
Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.H3 other text : device not returned to the manufacturer.
 
Event Description
Patient with mdm insert dislocated at home approx 2 months post op and had to have his hip revised.Update as per sales rep: the head came dislocated (disassociated) from the poly.It was still on the stem.There was a fall involved.Right side.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESTORATION ADM X3 INS 28/52
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18339950
MDR Text Key330667096
Report Number0002249697-2023-01567
Device Sequence Number1
Product Code MEH
UDI-Device Identifier07613327315967
UDI-Public07613327315967
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182468
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number7236-2-852
Device Lot Number88041401
Was Device Available for Evaluation? No
Date Manufacturer Received11/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age50 YR
Patient SexMale
-
-