MEDOS INTERNATIONAL SàRL COUPLER AC FFL 19; ACCESSORIES, PHOTOGRAPHIC, FOR ENDOSCOPE (EXCLUDE LIGHT SOURCES) Back to Search Results
Catalog Number 242437
Device Problems Moisture Damage (1405); Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/20/2023
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: e3: reporter is a j&j sales representative.H4: the device manufacture date is unknown.(b)(4) investigation summary the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: ¿ moisture inside optical system.Per service reports, this complaint can be confirmed.The device was scrapped as system is non-repairable.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.The user error was identified as the root cause for the device failure during the service evaluation.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
It was reported by the sales rep on (b)(6)2023 that when the coupler ac ffl 19 device was being washed outside of surgery, it appeared that there was water inside the lens.There was no patient involvement.During in-house engineering evaluation it was determined that there was moisture inside the optical system.There were no reported adverse patient consequences.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COUPLER AC FFL 19
Type of Device
ACCESSORIES, PHOTOGRAPHIC, FOR ENDOSCOPE (EXCLUDE LIGHT SOURCES)
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key18340148
MDR Text Key330781430
Report Number1221934-2023-04601
Device Sequence Number1
Product Code FEM
UDI-Device Identifier10886705028610
UDI-Public10886705028610
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number242437
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2023
Date Manufacturer Received12/13/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown