Additional information: h3, h4, h6, h10.H10: the actual device was not available; however, two (2) retained samples were evaluated.A visual inspection was performed which did not identify any abnormalities that could have contributed to the reported condition.Functional push-pull, underwater leak, and air passage testing were performed; no leaks or disconnections were identified, and no blockages were found.Traction testing was performed with no issues noted.The reported condition was not verified on the retained samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|