• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION TUR IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION TUR IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL Back to Search Results
Catalog Number VMC4005
Device Problems Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2023
Event Type  malfunction  
Event Description
It was reported that one of the tubes of an uromatic y type tur set separated.The issue was identified during priming.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
E1: initial reporter first name: (b)(6).E1: initial reporter last name: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Additional information: h3, h4, h6, h10.H10: the actual device was not available; however, two (2) retained samples were evaluated.A visual inspection was performed which did not identify any abnormalities that could have contributed to the reported condition.Functional push-pull, underwater leak, and air passage testing were performed; no leaks or disconnections were identified, and no blockages were found.Traction testing was performed with no issues noted.The reported condition was not verified on the retained samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUR IRRIGATION SET
Type of Device
SYSTEM, IRRIGATION, UROLOGICAL
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MALTA
a47 industrial estate
marsa
MT  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18340880
MDR Text Key330987807
Report Number1416980-2023-06573
Device Sequence Number1
Product Code LJH
UDI-Device Identifier05413760232324
UDI-Public(01)05413760232324
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberVMC4005
Device Lot Number22K23V492
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/12/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NA.
-
-