|
Model Number AIA-900 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/08/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A field service engineer (fse) was at the customer¿s site to address the reported event.Fse confirmed the complaint by reviewing the error logs.The fse powered off and on, moved the wash syringe by hand to test functionality.Fse replaced the washer syringe assembly, cable (lqc board to cn2) and (s100) pip electric board.Fse repaired and validated the analyzer by successfully performing quality control run without error and within acceptable range.No further action required by field service.The aia-900 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(6).There were no similar complaints identified during the searched period.The aia-900 operator's manual under section 12 - flags and error messages states the following: (4251) wash-syr home detect error.Cause: the home sensor s100 failed to be activated after the wash syringe moved toward the home position.No further operation will take place.Action: remove the impediment or other cause of the error.Check s100 and pm100 for a possible malfunction.The most probable cause of the reported event was due to the failures of the washer syringe assembly, cable, and (s100) pip electric board.
|
|
Event Description
|
A customer reported error message ¿4251 wash-syr home detect error¿ during startup on the aia-900 analyzer.The customer powered the analyzer off and on.Technical support specialist (tss) instructed the customer to observe the bf probes and the tips were seated properly and performed an all-home-set, the error persists.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event which resulted in a delayed reporting of patient samples for beta human chorionic gonadotropin (bhcg).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
|
Manufacturer Narrative
|
The pip electric board, washer syringe assembly, and cable (lqc board to cn2) were returned to the instrument service center for investigation.A visual inspection was performed on the returned parts and found the pip electric board was damaged and does not met specifications, therefore leading to component failure.No visual damage for the washer syringe assembly and cable was observed.Functional testing was performed using a digital multimeter and no open circuits were found.The washer syringe assembly and cable passed within specifications and no device problem was found.The most probable cause of the reported event was due to the faulty pip electric board.
|
|
Search Alerts/Recalls
|
|
|