• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC QDOT MICRO¿ CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC QDOT MICRO¿ CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Catalog Number D139505
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/11/2023
Event Type  malfunction  
Event Description
It was reported that a patient underwent a cardiac ablation procedure with a qdot micro¿ catheter for which biosense webster¿s product analysis lab (pal) identified a hole in the pebax.It was initially reported by the customer there was a force sensor error and blood in the tip of catheter.No consequences for the patient reported.To procedure was completed by the physician when they used another product the same type.The customer¿s reported issue of sensor error is not considered to be mdr reportable since the potential risk that it could cause or contribute to a serious injury or death to the operator or patient is remote.On 23-nov-2023, the bwi pal revealed that a visual inspection of the returned device found a hole in the pebax.This finding was reviewed and assessed the issue of a ¿hole¿ in the pebax as an mdr reportable malfunction since the integrity of the device has been compromised.
 
Manufacturer Narrative
Device evaluation details: the device was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.A visual inspection and screening test of the returned device were performed in accordance with bwi procedures.Visual analysis revealed a hole in the pebax and foreign reddish material was observed inside it.A screening test was performed, and the device was visualized and recognized correctly; no force issues were observed, and since the sensor values were found within specifications, the event described by the customer could be related to the reddish material inside the pebax, however, the root cause of the damage on the pebax could not be determined; it was concluded that it occurred outside the bwi manufacturing facilities.A manufacturing record evaluation was performed for the finished device 31055915l, and no internal action was found during the review.The issues reported by the customer were confirmed.It should be noted that product failure is multifactorial.The instructions for use contain the following recommendations: the force sensor of the catheter is disconnected.If the problem persists, replace the catheter cable or the catheter.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QDOT MICRO¿ CATHETER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18344918
MDR Text Key331087275
Report Number2029046-2023-02989
Device Sequence Number1
Product Code OAE
UDI-Device Identifier10846835016758
UDI-Public10846835016758
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
P210027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD139505
Device Lot Number31055915L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2023
Date Manufacturer Received11/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-