Catalog Number 07236450190 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457); Non Reproducible Results (4029)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/29/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The analyzer serial number is (b)(6).The customer suspects an interfering factor in the patient's sample.The investigation is ongoing.
|
|
Event Description
|
There was an allegation of questionable nh3l ammonia results for 1 patient sample on a cobas 8000 c702 module.The initial nh3 result was 135.1 umol/l.The customer was prompted to repeat the sample as there are internal protocols to rerun all samples that have a nh3 result of >100 umol/l.The sample was repeated twice and the results were 338.6 umol/l and 436.8 umol/l.
|
|
Manufacturer Narrative
|
The investigation found that the patient was taking erwinase.Per erwinase labeling, the drug can lead to an increase in ammonia in the blood in 1% of patients.The root cause of the event was found to be consistent with erwinase interference.No product problem was identified.
|
|
Search Alerts/Recalls
|