Catalog Number P200D/NJJP
Device Problems Break (1069); Fracture (1260); Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Event Description
It was reported that the device needed inspection and repair due to damage and leakage of oxygen hose.The event occurred before patient use.
 
Manufacturer Narrative
B3: date of event and d4: udi number are unknown; no information has been provided to date.One device was received.No physical damage was found on the device.Per functional testing, there was a leak from the gas hose.The complaint was confirmed.The root cause was due to deterioration of the gas hose.It was unknown what caused the condition.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.The gas hose was replaced.Replaced alarm gauge bezel, light emitting diode (led) printed circuit board (pcb) set, and switch cover due to damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC PARAPAC W/ALARMS
Type of Device
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL LTD.
enterprise way
luton LU3 4 BU
UK  LU3 4BU
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18352439
MDR Text Key331067866
Report Number9611178-2023-00007
Device Sequence Number1
Product Code BTL
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberP200D/NJJP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2023
Date Manufacturer Received11/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/22/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse