ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problems Failure to Advance (2524); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented for initial implant.During the procedure, the left ventricular lead failed to advance over the guidewire.It was also found that the lead was kinked.The lead was not used and replaced.There were no patient consequences.
 
Manufacturer Narrative
The reported event of ¿lead while advancing got kinked and wire not getting passed¿ was not confirmed.Final analysis found that as received, a complete lead was returned in one piece.The guidewire used in the field was not returned with the lead for analysis.The guidewire insertion test was performed, and the guidewire could be fully inserted and removed without difficulties.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual and x-ray examinations of the lead did not find any anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18352682
MDR Text Key330847026
Report Number2017865-2023-95026
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberA000137366
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-