MAUDE Adverse Event Report: COCHLEAR LIMITED NUCLEUS CI622 COCHLEAR IMPLANT WITH SLIM STRAIGHT ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR LIMITED NUCLEUS CI622 COCHLEAR IMPLANT WITH SLIM STRAIGHT ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number CI622
Device Problems Circuit Failure (1089); High impedance (1291); Therapeutic or Diagnostic Output Failure (3023)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 07/07/2022
Event Type  malfunction  
Manufacturer Narrative
This report is submitted on december 19, 2023.
 
Event Description
Per the clinic, the patient experienced poor performance, open circuit electrodes and high impedance with the internal device.Troubleshooting at the clinic did not resolve the issue and the clinic requested a cochlear specialist to verify the device function.Subsequently, the device was explanted on (b)(6) 2023 and the patient was reimplanted with another cochlear device during the same surgery.
 
Manufacturer Narrative
Device analysis indicated device failure.Device analysis report attached.This report is submitted on february 19, 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI622 COCHLEAR IMPLANT WITH SLIM STRAIGHT ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LIMITED
1 university avenue
macquarie university
nsw 2109
AS  2109
Manufacturer (Section G)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS   2109
Manufacturer Contact
mastura ruhanet
unit ug-1, vertical podium
no.8 jalan kerinchi
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key18352947
MDR Text Key330810020
Report Number6000034-2023-04200
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502036573
UDI-Public(01)09321502036573(11)210916(17)230915
Combination Product (y/n)N
Reporter Country CodeEI
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/15/2023
Device Model NumberCI622
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received01/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age18 MO
Patient SexMale
-
-