The device was returned and evaluated, and the issue was determined to be due to damage on charged coupled device unit.In addition to the malfunction documented in b5, the additional evaluation findings are as follows: due to wear of angle wire, bending angle in up direction does not meet the standard value, due to damage on charged coupled device unit, the image disappears during angulation in upward/downward angulation directions, suction connector is dirty due to water leakage, air/water leakage from the distal end, the image is distorted during angle operation but returns to normal (the subject can be recognized).The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
The customer reported to olympus, the evis exera iii bronchovideoscope had a distorted image.The device was returned for evaluation.During the device evaluation, it was found that there was no image.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|