• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. FEMORAL HEAD/NECK IMPACTOR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. FEMORAL HEAD/NECK IMPACTOR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Catalog Number 71360093
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/27/2023
Event Type  malfunction  
Event Description
It was reported that during a thr surgery, in one (1) femoral head/neck impactor, the black tip impactor fractured after impacting head.No pieces fell into patient.The procedure was resumed, without any delay, using a s+n back-up device.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
Internal complaint number: (b)(4).
 
Manufacturer Narrative
H3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history over 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD/NECK IMPACTOR
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18357771
MDR Text Key331209485
Report Number1020279-2023-02524
Device Sequence Number1
Product Code LZO
UDI-Device Identifier03596010619310
UDI-Public03596010619310
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number71360093
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/10/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage A
Patient Sequence Number1
-
-