Catalog Number EMAX2PLUS |
Device Problems
Environmental Particulates (2930); Fail-Safe Did Not Operate (4046)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.The device was evaluated and it was determined that the reported condition was not confirmed.Therefore, the assignable root cause was not determined.However, during evaluation, it was determined that the device cord and flexi circuit were damaged, was loose, had low power, made excessive noise, the bearing was worn, heated up, run in the locked position, was vibrating, had component damaged and the speed could not be changed.The assignable root cause was traced to component failure due to normal wear.(b)(4).
|
|
Event Description
|
It was reported from japan that during service and evaluation, it was determined that the motor device cord and flexi circuit were damaged, the hose was scratched, the connector and set screw were loose, was spinning in safe mode, had low power, made excessive noise, the bearing was worn, heated up, run in the locked position, was vibrating, had component damaged and the speed could not be changed.It was further determined that the device failed pretests for visual assessments, loctite assessment, safety assessment, rotational speed assessments, noise assessment, handpiece temperature assessment and hand control assessment.It was noted in the service order that the device was emitting smoke this event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|