Catalog Number EMAX2PLUS
Device Problems Environmental Particulates (2930); Fail-Safe Did Not Operate (4046)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2023
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.The device was evaluated and it was determined that the reported condition was not confirmed.Therefore, the assignable root cause was not determined.However, during evaluation, it was determined that the device cord and flexi circuit were damaged, was loose, had low power, made excessive noise, the bearing was worn, heated up, run in the locked position, was vibrating, had component damaged and the speed could not be changed.The assignable root cause was traced to component failure due to normal wear.(b)(4).
 
Event Description
It was reported from japan that during service and evaluation, it was determined that the motor device cord and flexi circuit were damaged, the hose was scratched, the connector and set screw were loose, was spinning in safe mode, had low power, made excessive noise, the bearing was worn, heated up, run in the locked position, was vibrating, had component damaged and the speed could not be changed.It was further determined that the device failed pretests for visual assessments, loctite assessment, safety assessment, rotational speed assessments, noise assessment, handpiece temperature assessment and hand control assessment.It was noted in the service order that the device was emitting smoke this event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMAX 2 PLUS MOTOR
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key18357989
MDR Text Key330900603
Report Number1045834-2023-01731
Device Sequence Number1
Product Code GEY
UDI-Device Identifier00845384002055
UDI-Public00845384002055
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K080802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEMAX2PLUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2023
Date Manufacturer Received12/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/17/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-