Catalog Number 101011090
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Infection (1930)
Event Date 12/07/2023
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that infected hip prostheses were removed and a prostalac was placed uneventfully after a thorough debridement and i&d were performed.The head and dm poly liner that were removed had been placed on (b)(3) 2023 due to dislocation.There was no surgical delay and no patient harm occurred.Doi: (b)(3) 2023 (other implants).Doi: (b)(3) 2023 (the head and dm poly liner).Dor: (b)(3) 2023.Affected side: left hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIS COLLARED STD SIZE 9
Type of Device
HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18358904
MDR Text Key330907103
Report Number1818910-2023-25630
Device Sequence Number1
Product Code MEH
UDI-Device Identifier10603295380498
UDI-Public10603295380498
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K202472
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number101011090
Device Lot NumberJ6310X
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/30/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BI MENTUM ALTRX LNR 51 28.; DLT TS CER HD 12/14 28MM +8.5.; PINN CAN BONE SCREW 6.5MMX25MM.; PINN SECTOR W/GRIPTION 60MM.; PINNACLE DM LINER 60_51.
Patient Outcome(s) Required Intervention;
Patient Age66 YR
Patient SexMale
Patient Weight95 KG
-
-